Appendix B Data Accuracy
Accuracy of data on the database varies greatly and depends on the
measurement methods used.
Table 2 shows the accuracy of measurement of depth, temperature
and salinity using different gear methods. The gear methods are
listed down the side of the table and the physical quantities are
listed along the top. Some quantities may be measured using different
techniques for the same gear type (e.g. In a messenger bottle cast
depth may be measured in three different ways). The body of the table
contains accuracy of measurement and any variations in accuracy due
to the measurement techniques used.
Note that according to Intergovernmental Oceanographic Commission
Standards salinity is defined against a single reference point (S =
35), which has the same electrical conductivity as a reference
potassium chloride solution with a temperature of 15 oC at
atmospheric pressure, and is no longer measured in parts per thousand
(from Joint Panel on Oceanographic Tables and Standards 1991).
|
Method
|
Physical Quantity
|
|
Depth
|
Temperature(oC)
|
Salinity
|
|
Bottle Casts
|
|
- Usually
±0.02oC, can be as good as ±0.01oC
|
- Autolab salinometer
±0.010d
-
Guildline Autosal 8400 salinometer
±0.002e
|
|
CTD Probe
|
±0.15% (<2m at 2000m)e
|
±0.005oCe
|
±0.005 after calibration against water samplese, ±0.005
otherwisef
|
|
MOCNESS CTD
|
±5me
|
0.1e
|
Not yet tested.
|
|
MBTg
|
±1%
|
±0.5
|
|
|
XBTe
Mechanical recording
Electronic recording
|
±2% (a few metres)
±2% (a few metres)
|
±0.2
±0.1
|
|
Table 3: Accuracy of depth, temperature, and
salinity data acquired using various methods.
a Estimate only. If the wire angle is less than 5o the length of
wire can be used as a measure of depth.
b Estimate based on deviation from correct depth at maximum wire
angle of 20o, taking no account of variations in
subsurface currents.
c 0.01oC given in Sverdrup et al (1942) but thermometer
performance may change with time.
d Anthony Cole, NIWA, pers. Comm.
e Manufacturer's Manual
f This is the largest difference
between CTD
and rosette salinities that has been found from calibration against
bottle samples at Greta Point.
g From a set of locally-produced instructions on calibrating the
MBT.